[AliLO Nulling]:
Press [System]→[AliLO Nulling] to enable the LO
nulling alignment function. The user can easily apply this function when
required. Alignment results will be saved in the internal memory of the
instrument.
· [Date]:
Press [System]→ [Date/Time] to set the date and time.
· [Display setting]:
Press [System]→ [Display] to set the display mode,
including relevant settings of [Default], [Black & White], [Night
Vision] and [Brightness].
·[GPS]:
Press [System]→ [GPS] and soft menus related to GPS will pop
up, including [GPS Off On], [GPS Info] and [Reset]. You can enable
or disable the GPS function, view GPS details and reset GPS.
· [Special note]: The GPS positioning function is optional.
· [Internet setting]:
Press [System]→ [LAN] to set the network setting of
the spectrum analyzer, including [Prev], [Next], [IP], [Mask] and
[Gateway].
· [Freq Ref Int Ext]:
Press [System]→[Freq Ref Int Ext] to select the
internal or external frequency reference according to the needs.
· [Ref Output Off On]:
Press [System]→[Ref Output Off On] to enable or
disable the interval reference according to the needs in the internal
reference mode.
· [Special note]: The external reference frequency must be
10MHz±100Hz, and the amplitude must be 0dBm (range: -2dBm to
+10dBm). The external reference frequency must be applied through
the “10MHz reference input” of the cover.