6) Press [Measure]→ [Power Display Curr Max], and enable the “Current” option. Set the current power of each
channel.
7) Press [Measure]→ [Color Code Single Dual] and enable the dual color mode.
8) Press [Measure]→ [Orientation Vert Hori] and enable the horizontal mode.
Fig. 8-2 Schematic Diagram of Frequency Scanner
3
List scanner
Below is an example of the Frequency Scanner mode, mainly involving the following procedures.
1) Press [Sweep]→ [MScan] → [Edit List] and edit the channel list to be swept. The information of each channel
can be set by setting the signal standard and channel number or setting the frequency and bandwidth in the list.
2) Press [Sweep]→ [MScan] →[Number of Channels], and set the number of measured channels. At most 20
channels can be measured at the same time.
3) Press [Sweep]→ [Display Graph Table] and enable the graph display mode.