frequency and current spectrometer scan frequency.
·[Special Notice]:
The parameter range is -300MHz - 300MHz with default
value of 0Hz, which is effective when the tracking mode is activated. The
frequency offset must be set to ensure that the generator can reach the
maximum frequency at 100kHz.
·[Normalize Off On]:
press [Measure]→[More 1 of 2] →[Generator] →
[Trans Meas] → [Normalize Off On], to activate or deactivate the
normalized measurement.
·[Ref Level 0.0dB]:
press [Measure]→[More 1 of 2] →[Generator] →
[Trans Meas] →[Ref Level 0.0dB], to activate the normalized and adjust
the ref level so as to adjust the vertical position of trace on screen.
·[Special Notice]: The parameter range is -200dB to 200dB with
default value of 0dB and step size of 1dB, which is effective when the
tracking mode and normalized are activated.
·[Ref Position 0]:
press [Measure] →[More 1 of 2] →[Generator] →
[Trans Meas] →[Ref Position 0], to activate the normalized and adjust the
ref position so as to adjust the vertical position of normalized ref level on
screen.
·[Special Notice]:
The parameter range is 0 - 10 with default value of 5
and step size of 1, which is effective when the tracking mode is activated.
·[Scale/Div 10.0dB]:
press [Measure]→[More 1 of 2]→[Generator]→
[Trans Meas] →[Scale/Div 10.0dB], to activate the normalized and adjust
the scale/div so as to adjust the Y-axis precision of trace on screen.
·[Special Notice]:
The parameter range is 0~10 with default value of 5
and step size of 1, which is effective when the tracking mode is activated.
·[Store Mem Trace]:
press [Measure]→[More 1 of 2]→[Generator]→
[Trans Meas]→[Store Mem Trace]. It is allowed to save the data of current
mem trace.
·[Mem Trace Off On]
: press [Measure]→[More 1 of 2]→[Generator]→
[Trans Meas]→[Mem Trace Off On]. Set whether the mem trace is
displayed or not.