7 Sweep menu
·[Sweep Time Auto Man]:
Press [Sweep]→[Sweep Time Auto Man] to
switch the sweep time between the auto and manual modes. The current
state is underlined. In the auto mode, the minimum sweep time will be set
automatically by the integrated test instrument based on the current state
and displayed on the screen. In the manual mode, you can enter the
sweep time with the number keys and select the time unit with
corresponding soft keys. When RBW and VBW are changed in the auto
mode, the sweeping speed will change accordingly. The larger the RBW
and VBW values are, the higher the sweeping speed is; and vice versa.
Based on the compliance with minimum sweep time constraints, the
maximum sweep time of S3302 series spectrum analyzer can be set as
max. 800s under the non-zero span and max. 600s under zero span.
·[Sweep Cont Single]:
Press [Sweep]→[Sweep Cont Single]. The sweep
type is decisive to the sweeping mode of the integrated test instrument and
the time to stop sweeping and start holding. Two options are available in
the interference analyzer mode: continuous and single.
·[Sweep once]:
Press [Sweep]→[Sweep once] to sweep once again.