Reference Manual
00809-0100-4663, Rev BA
January 2010
Rosemount 8732
E-10
Model 10D1419 Sensor
to Rosemount 8732
Transmitter
Connect coil drive and electrode cables as shown in Figure E-7.
Figure E-7. Wiring Diagram for
Fischer and Porter Sensor
Model 10D1419 and Rosemount
8732
Table E-8. Fischer and Porter
Model 10D1419 Sensor Wiring
Connections
ROSEMOUNT 8732 TRANSMITTER
3
2
1
18
L1
L2
16
17
Electrode Connections
Coil Connections
Rosemount 8732
Fischer and Porter Model 10D1419 Sensors
1
L1
2
L2
3
17
3
18
1
19
2
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.