Rosemount 8732
ROSEMOUNT SENSORS
Rosemount
8705/8707/8711/8721
Sensors to Rosemount
8732 Transmitter
Connect coil drive and electrode cables as shown in Figure .
Figure E-1. Wiring Diagram to a
Rosemount 8732 Transmitter
Table E-2. Rosemount
8705/8707/8711/8721 Sensor
Wiring Connections
Rosemount 8732 Transmitters
Rosemount 8705/8707/8711/8721 Sensors
1
1
2
2
17
17
18
18
19
19
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.