Reference Manual
00809-0100-4663, Rev BA
January 2010
6-3
Rosemount 8732
DIAGNOSTIC
MESSAGES
Problems in the magnetic flowmeter system are usually indicated by incorrect
output readings from the system, error messages, or failed tests. Consider all
sources in identifying a problem in your system.
Table 6-1. Rosemount 8732 Basic Diagnostic Messages
Message
Potential Cause
Corrective Action
“Fieldbus Not
Communicating”
Fieldbus segment is disconnected
Connect the fieldbus segment
Fieldbus segment power missing
Verify the segment fieldbus voltage
Electronics failure
Replace electronics
“Sensor Processor Not
Communicating”
Transmitter input power (AC/DC) is
not connected
Connect the input power. If the LCD displays a message, the input power is
applied
Electronics failure
Replace electronics
“Empty Pipe”
Empty Pipe
None - message will clear when pipe is full
Wiring Error
Check that wiring matches appropriate wiring diagrams - see Appendix E:
Universal Sensor Wiring Diagrams
Electrode Error
Perform sensor tests C and D (see Table 6-5 on page 6-8)
Conductivity less than 5
microsiemens per cm
Increase Conductivity to greater than or equal to 5 microsiemens per cm
Intermittent Diagnostic
Adjust tuning of Empty Pipe parameters
“Coil Open Circuit”
Improper wiring
Check coil drive wiring and sensor coils
Perform sensor test A - Sensor Coil
Other manufacturer’s sensor
Change coil current to 75 mA
Perform a Universal Auto Trim to select the proper coil current
Circuit Board Failure
Replace Rosemount 8732 Electronics
Coil Circuit OPEN Fuse
Return to factory for fuse replacement
“Auto Zero Failure”
(Cycle power to clear
messages, no changes
were made)
Flow is not set to zero
Force flow to zero, perform autozero
Unshielded cable in use
Change wire to shielded cable
Moisture problems
See moisture problems in “Accuracy Section”
Empty pipe is present
Fill sensor with process fluid
“Universal Trim Failure”
No flow in pipe while performing
Universal Auto Trim
Establish a known flow in sensor, and perform Universal Auto-Trim
calibration
Wiring error
Check that wiring matches appropriate wiring diagrams - see “Universal
Sensor Wiring Diagrams” on page E-1
Flow rate is changing in pipe while
performing Universal Auto-Trim
routine
Establish a constant flow in sensor, and perform Universal Auto-Trim
calibration
Flow rate through sensor is
significantly different than value
entered during Universal Auto-Trim
routine
Verify flow in sensor and perform Universal Auto-Trim calibration
Incorrect calibration number
entered into transmitter for
Universal Auto-Trim routine
Replace sensor calibration number with 1000005010000001
Wrong sensor size selected
Correct sensor size setting - See “Line Size” on page 3-9
Sensor failure
Perform sensor tests C and D (see Table 6-5 on page 6-8)
“Electronics Failure”
Electronics self check failure
Replace Electronics
“Electronics Temp Fail”
Ambient temperature exceeded the
electronics temperature limits
Move transmitter to a location with an ambient temperature range of -40 to
165 °F (-40 to 74 °C)
“Reverse Flow”
Electrode or coil wires reverse
Verify wiring between sensor and transmitter
Flow is reverse
Turn ON Reverse Flow Enable to read flow
Sensor installed backwards
Re-install sensor correctly, or switch either the electrode wires (18 and 19)
or the coil wires (1 and 2)
“Flow Rate > Sensor Limit” Flow rate is greater than 43 ft/sec
Lower flow velocity, increase pipe diameter
Improper wiring
Check coil drive wiring and sensor coils
Perform sensor test A - Sensor Coil (see Table 6-5 on page 6-8)
“Digital Trim Failure”
(Cycle power to clear
messages, no changes
were made)
The calibrator (8714B/C/D) is not
connected properly
Review calibrator connections
Incorrect calibration number
entered into transmitter
Replace sensor calibration number with 1000005010000001
Calibrator is not set to 30 FPS
Change calibrator setting to 30 FPS
Bad calibrator
Replace calibrator
Summary of Contents for 8732 Series
Page 2: ......
Page 4: ......
Page 10: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 TOC 6 ...
Page 26: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 2 14 ...
Page 38: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 3 12 ...
Page 92: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 A 8 ...
Page 186: ...Reference Manual 00809 0100 4663 Rev BA January 2010 Rosemount 8732 H 8 ...
Page 189: ......