Remote control commands
R&S
®
RTM3000
782
User Manual 1335.9090.02 ─ 09
WGENerator:ARBitrary[:FILE]:NAME
............................................................................... 782
WGENerator:ARBitrary[:FILE]:OPEN
............................................................................... 782
........................................................................................783
WGENerator:ARBitrary:SOURce
<Source>
Selects the input channel for an arbitrary function to be generated.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4
*RST:
CH1
WGENerator:ARBitrary:RANGe:START
<StartTime>
Sets the start time of the copied waveform part.
Parameters:
<StartTime>
Default unit: s
WGENerator:ARBitrary:RANGe:STOP
<StopTime>
Sets the end time of the copied waveform part.
Parameters:
<StopTime>
Default unit: s
WGENerator:ARBitrary:UPDate
Loads the waveform from the selected signal source (
).
Usage:
Event
WGENerator:ARBitrary[:FILE]:NAME
<FilePath>
Sets the file path and the file for an arbitrary waveform to be loaded.
Parameters:
<FilePath>
string
*RST:
"
WGENerator:ARBitrary[:FILE]:OPEN
Loads the arbitrary waveform, that is selected with the
command.
Usage:
Event
Signal generation (option R&S