Remote control commands
R&S
®
RTM3000
626
User Manual 1335.9090.02 ─ 09
Suffix:
<b>
.
1..4
<n>
*
Selects the frame.
Return values:
<AddressValue>
Decimal value
Range:
0 to 2047
Increment: 1
Usage:
Query only
BUS<b>:I2C:FRAMe<n>:ADEVice?
Returns the decimal address value of the indicated frame
without
R/W bit.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame.
Return values:
<SlaveAddress>
Decimal value
Range:
0 to 1023
Increment: 1
Usage:
Query only
BUS<b>:I2C:FRAMe<n>:AMODe?
Returns the address length.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame.
Return values:
<AddressMode>
BIT7 | BIT10
Usage:
Query only
BUS<b>:I2C:FRAMe<n>:ASTart?
Returns the start time of the address for the indicated frame.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame.
Serial bus analysis