Applications
R&S
®
RTM3000
197
User Manual 1335.9090.02 ─ 09
The simple peak setup needs only the setting of the minimum level for peak detection.
More complex criteria can be defined in the advanced setup.
Min. Level
Defines the minimum absolute level for peak detection. Only peaks that exceed the
minimum level are detected, listed in the "Peak List", and marked in the spectrum dia-
gram.
Remote command:
Advanced Peak Setup
Enables a more precise peak detection.
For a peak to be detected, the
,
have to be within
the defined limits.
Remote command:
Max. Width
Defines the maximum width that a peak can have for it to be detected.
Remote command:
Excursion
Defines a minimum level difference between two subsequent peaks that has to be kept
for the peak to be detected. To avoid identifying noise peaks, enter a peak excursion
value that is higher than the noise levels.
Remote command:
SPECtrum:MARKer:SETup:EXCursion
Spectrum analysis and spectrogram (option R&S
RTM-K37)