Remote control commands
R&S
®
RTM3000
517
User Manual 1335.9090.02 ─ 09
PPCount | NPCount
Counts positive or negative pulses.
RECount | FECount
Counts the number of rising or falling edges.
HIGH | LOW
Mean value of the high or low level of a square wave.
AMPLitude
Amplitude of a square wave.
MEAN
Mean voltage value of the complete displayed waveform of the
selected channel.
RMS
RMS (root mean square) voltage value of the complete dis-
played waveform of the selected channel.
RTIMe | FTIMe
Rise or falling time of the left-most rising or falling edge within
the displayed section of the waveform of the selected channel.
The reference level for this measurement is set with
SRRise | SRFall
Slew rate of the left-most rising or falling edge within the dis-
played section of the waveform of the selected channel. The ref-
erence level for this measurement is set with
PDCycle | NDCycle
Positive or negative duty cycle.
PPWidth | NPWidth
Width of positive or negative pulses.
CYCMean
Mean voltage value of the left-most signal period of the wave-
form of the selected channel.
CYCRms
RMS (root mean square) voltage value of the left-most signal
period of the waveform of the selected channel.
STDDev
Standard deviation of the waveform.
CYCStddev
Standard deviation of one cycle, usually of the first, left-most sig-
nal period.
DELay
Time difference between two edges of the same or different
waveforms.
The edges are defined with
MEASurement<m>:DELay:DIRection
.
Measurements