Remote control commands
R&S
®
RTM3000
611
User Manual 1335.9090.02 ─ 09
BUS<b>:SSPI:CLOCk:POLarity
<Polarity>
Selects if data is stored with the rising or falling slope of the clock. The slope marks the
begin of a new bit.
Suffix:
<b>
.
1..4
Parameters:
<Polarity>
POSitive | NEGative
POSitive: rising slope
NEGative: falling slope
*RST:
POSitive
BUS<b>:SSPI:DATA:SOURce
<Source>
BUS<b>:SSPI:MOSI:SOURce
<MosiSource>
Selects the input channel of the MOSI / MISO line.
Suffix:
<b>
.
1..4
Parameters:
<MosiSource>
CH1 | CH2 | CH3 | CH4 | D0..D15
*RST:
CH1
BUS<b>:SSPI:MISO:SOURce
<MisoSource>
Selects the input channel of the optional MISO line.
Suffix:
<b>
.
1 | 3. Bus 2 or 4 is not available if the MISO source is used on
bus 1 or 3, respectively.
Parameters:
<MisoSource>
CH1 | CH2 | CH3 | CH4 | NONE | D0..D15
*RST:
NONE
BUS<b>:SSPI:DATA:POLarity
<Polarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1) on the
data line.
Suffix:
<b>
.
1..4
Parameters:
<Polarity>
POSitive | NEGative
POSitive = high active
NEGative = low active
*RST:
POSitive
Serial bus analysis