Remote control commands
R&S
®
RTM3000
460
User Manual 1335.9090.02 ─ 09
17.5
Trigger
.........................................................................................460
17.5.1
General trigger settings
This section describes general trigger commands that are independent of the trigger
type.
.............................................................................................461
...............................................................................................462
TRIGger:A:MODE
<TriggerMode>
Sets the trigger mode. The trigger mode determines the behavior of the instrument if
no trigger occurs.
Parameters:
<TriggerMode>
AUTO | NORMal
AUTO
The instrument triggers repeatedly after a time interval if the trig-
ger conditions are not fulfilled. If a real trigger occurs, it takes
precedence.
NORMal
The instrument acquires a waveform only if a trigger occurs.
*RST:
AUTO
TRIGger:A:SOURce
<Source>
Sets the trigger source for the selected A trigger type.
Parameters:
<Source>
CH1 | CH2 | CH3 | CH4 | EXTernanalog | LINE | SBUS1 |
SBUS2 | D0..D15
Trigger