Remote control commands
R&S
®
RTM3000
560
User Manual 1335.9090.02 ─ 09
Example:
DVM2:SOUR CH2
DVM2:TYPE DCRMs
DVM2:RES?
<-- 7.089E-01
An RMS measurement is performed on measurement place 2,
on channel 2. The result is 708.9
mV.
Usage:
Query only
DVM<m>:RESult[:ACTual]:STATus?
Returns the result value and the status of the result.
The status is the decimal representation of a 4-bit register value:
●
Bit 0 = 1: result is valid
●
Bit 1 = 1: no result available
●
Bit 2 = 1: clipping occurs
●
Bit 3 = 1: no period found
Suffix:
<m>
.
1..4
Sets the number of the meter measurement.
Return values:
<CurrentValue>
Measured value
<Status>
Decimal status value
Example:
DVM:SOUR CH1
DVM:TYPE MEAN
DVM:RES:STAT?
<-- 4.968E-01,5
The result value of the mean measurement on channel 1 is
496.1
mV. The result status is 5 (decimal) = 0101 (binary). That
means, the result is valid (bit 0 = 1), and the signal is clipped by
the limits of the ADC range (bit 3 = 1).
Usage:
Query only
17.8.7
Trigger counter
TCOunter:RESult[:ACTual]:FREQuency?
......................................................................... 561
TCOunter:RESult[:ACTual]:PERiod?
................................................................................ 561
TCOunter:ENABle
<Enable>
Enables or disables the trigger counter measurements.