![Roche OMNI C Instructions For Use Manual Download Page 108](http://html1.mh-extra.com/html/roche/omni-c/omni-c_instructions-for-use-manual_1472343108.webp)
6 Maintenance
6-14
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
6.6
Unscheduled
6.6.1 Cleaning the bottle compartment
•
Open the bottle compartment cover. The bottle exchange image appears on the display
(see Fig. 8).
•
Open the docking mechanisms and remove all bottles.
•
Clean the bottle compartment with a towel soaked in disinfectant (deproteinizer for
example).
•
Re-insert the bottles (see section
Using the empty C1 calibration solution 1 bottle as W
waste container, page 6-
and
Installing the waste water container, page 6-11
).
TIP: Do not scan a barcode!
•
While in the "Analyzer" operating mode, press:
>
System > Tools > Fluid actions > Auto preparation routines
•
Select successively "Prepare Calibration Solution C1" and "Prepare Calibration Solution
C2".
Start the process with
.
6.6.2 Cleaning the sample drip tray and wash plate
The sample drip tray and the wash plate are parts of the sample port module.
Handle these parts with care. Danger of injury!
Always wear gloves! Danger of infection!
While in the "Analyzer" operating mode, press:
>
System > Wash and clean > Clean sample port module
Fig. 16
Sample drip tray
Wash plate
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...