![Roche OMNI C Instructions For Use Manual Download Page 122](http://html1.mh-extra.com/html/roche/omni-c/omni-c_instructions-for-use-manual_1472343122.webp)
7 Trouble shooting
7-2
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
It is possible to access the menu "System" if no automatic actions are currently performed.
When changing to the menu "System" during a "system stop", the "system stop" will not be
terminated, but reactivated when the menu "System" is closed again. If e.g. there is a change
from a "Temperature system stop" to the menu "System" and the menu "System" is closed
again, the analyzer will display the system stop "Temperature" again. In the case a system
calibration has been started in the menu "System", it can not be conducted immediately
because the "system stop" has to be removed first.
After all "system stops" have been removed, "system stop consequence actions" will be exe-
cuted. There is a common list of consequence actions, which impedes double actions. This
means that if e.g. "washing" was activated several times, it is then performed only once.
Possible consequence actions are:
•
Warm-up
•
Sample sensor calibration
•
Prepare solution C1
•
Prepare solution C2
•
Prepare Pack C3 solutions
•
Fill the AutoQC wash tube
•
Fill the FMS reservoir
•
Wash
•
Wash AutoQC
•
Conductivity calibration
•
System calibration
•
Waste container fill level measurement
•
Aspirate Mix1
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...