![Roche OMNI C Instructions For Use Manual Download Page 123](http://html1.mh-extra.com/html/roche/omni-c/omni-c_instructions-for-use-manual_1472343123.webp)
7 Trouble shooting
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
7-3
7.2
List of system stops
7.2.1 3001 Measuring chamber cover open
Cause
•
The measuring chamber cover is open
•
The measuring chamber cover sensor (hall sensor) is defective
•
The measuring chamber cover cable is defective
Remedy
•
Close measuring chamber cover
•
Check the measuring chamber cover sensor ("More functions > System > Test > Control
sensors > Monitoring sensors")
•
In case of recurrence, call customer service (check cable, change components if neces-
sary)!
Consequence actions
1. 1. After opening for longer than 5 seconds: Warm-up
2. Wash
3. After changing (an) electrode(s): system calibration
7.2.2 3002 Bottle compartment cover open
Cause
•
The bottle compartment is open
•
The bottle compartment cover micro switch is defective
Remedy
•
Close bottle compartment cover
•
Check to be sure that the bottles are inserted completely
•
Check the bottle compartment cover micro switch ("More functions > System > Test >
Control sensors > Monitoring sensors")
•
In case of recurrence, call customer service (bottle compartment cover micro switch
change if necessary)!
Consequence actions
•
C1 changed: prepare solution C1, fill FMS reservoir, conductivity calibration
•
C2 changed: prepare solution C2, fill FMS reservoir, conductivity calibration
•
C3 changed resp. C3 docking mechanism micro switch operated: prepare C3 solutions
•
Waste container changed resp. micro switch operated: Waste container fill level meas-
urement
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...