6 Maintenance
6-20
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
6.6.6 Cleaning the measurement chamber
•
Open the top cover and measurement chamber covers.
•
Remove all electrodes as described in section:
Replacement of the electrodes and
the MCon, page 6-6
.
•
Clean the measuring chamber with a towel soaked in disinfectant (deproteinizer for
example).
•
Re-insert the electrodes.
TIP: Do not scan a barcode!
6.6.7 Cleaning the screen
While in the "Analyzer" operating mode, press:
>
System > Wash & clean > Clean screen
The keys on the screen are deactivated for 30 seconds.
Clean only with a moist cloth (for example, one that is soaked with disinfectant).
Do not use sprays!
After 30 seconds, the display changes back into its active condition.
6.6.8 Surfaces
IMPORTANT: Do not attempt to decontaminate any part of the instrument before
shutting it down and unplugging it from the power source.
Before plugging the instrument back in and turning it on, always wait 15 minutes
to allow the disinfectant to evaporate – Danger of fire and explosion!
For safety reasons, only authorized customer service personnel may decontaminate
the power pack!
Regularly decontaminate all outside surfaces of the instrument, including all covers (for
example, printer cover, bottle compartment cover, top cover). For these tasks, use the
decontaminant in accordance with local regulations.
Very dirty surfaces should first be cleaned with swabs or paper towel of gauze that have been
soaked in distilled water. Spray all removable covers (top cover, bottle compartment cover)
with surface disinfectant and then wipe them down with swabs or paper towel of gauze.
Always respect the time necessary for proper effects.
IMPORTANT:
Never spray parts that cannot be removed or that are inside the instrument!
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...