54
SCT-1100 Weight Transmitter
6.2
Serial Outputs
The instrument has two bidirectional serial outputs which are ASCII code compatible with most printers, remote
displays, PCs and other devices. See
for connection information.
Transmission of data through the serial ports can be configured in the parameters
,
and
in
the setup mode. See
6.2.1
COM1 Serial Port
The COM1 serial port is bi-directional (half duplex) and uses RS-485 for transmitting data. It is mainly used to
connect PCs, PLCs and additional remote displays. The transmission speed may be selected in the setup as: 1200,
2400, 4800, 9600, 19200, 38400, 57600, 115200 baud (bit/sec).
On the same RS-485 line, it is possible to connect up to 32 devices (instruments, RS-485/RS-232 signal converter).
Considerations when making RS-485 connections:
• Use an
STP
(Shielded Twisted Pair) cable to make the connection (twisted and shielded pair(s) with single
shielding for each pair through aluminum band and total shielding through external shielding)
• Using 2x24 AWG duplex cable with external shielding, the RS-485 cable should not exceed 3937' (1200 m).
See
• With very long cables, cable capacity becomes a dominant factor in power consumption (normally near
50pF/m) and this increases when speed is increased; i.e. the maximum distance can’t be covered with the
maximum possible speed.
RS-485 Serial Address – In the
functioning mode it is possible to enable the RS-485 protocol and link its
address to the instrument; the instrument then repeats the weight if the address in the received string is the same
as set here
NOTE: The parameter displays only in the secondary mode, when
is selected as the functioning
mode and the
is set to
,
or
.
RS-485 protocol is disabled
RS-485 protocol is enabled; prompt to enter the RS-485 address displays,
in which
is the RS-485 address; enter the RS-485 address (00 to 99 in repeater mode);
Press the
or
to increases or decrease the digit and
select the digit to be modified;
when complete, confirm the address by pressing
Communication time out – When in the
functioning mode the communication time out can be set; Once the
alloted time has passed with no communication, the center segments of the LED display. Default 01.5 sec
NOTE: MAX=20.0 sec; MIN=00.5 sec
This parameter displays when
mode is selected in the secondary instrument.
Terminal Position
Indicator Serial Line
TX+/RX+
22A(+)
TX-/RX
23B(-)
Table 6-3. RS-485 Connections
Baud Rate (bit/sec)
Total Cable Capacity (pF)
1200
400000
2400
200000
4800
100000
9600
50000
19200
25000
38400
12000
57600
8000
115200
4000
Table 6-4. Cable Capacity
Parameter
Settings
Description
Table 6-2. Serial Menu Parameters and Settings (Continued)