Communications
59
Extended Weight Read Command With Date and Hour
[CC]REXD<CR LF>
Instrument response in the
or in the
mode: extended string (
).
Instrument response in the
mode: multi-scale string (
).
Weight Read Command
[CC]READ<CR LF>
Instrument response: standard string (
Weight Reading Command With Sensitivity Times 10
[CC]GR10<CR LF>
Instrument response: standard string (
Reading Command of MicroVolts Relative to the Weight
[CC]MVOL<CR LF>
Instrument response in
mode: standard string (
Response of the instrument in
and
mode: multi-scale string (
If the instrument is in the
mode (scale with independent channels) or in the
mode (scale with
dependent channels and digitally equalized) the weight value is read relative to the active channel; to read the
values of the other channels (if configured) switch to the desired channel. See
If the instrument is in the
mode (scale with independent channels) it is possible to read simultaneously
the values for all the configured channels.
Reading Command of Converter Points Relative to the Weight
[CC]RAZF<CR LF>
Instrument response in
mode: standard string (
Response of the instrument in
and
mode: multi-scale string (
If the instrument is in the
mode (scale with independent channels) or in the
mode (scale with
dependent channels and digitally equalized) the weight value is read relative to the active channel; to read the
values of the other channels (if configured) switch to the desired channel. See
If the instrument is in the
mode (scale with independent channels) it is possible to read simultaneously
the values for all the configured channels.
Tare Command
[CC]TARE<CR LF> or [CC]T<CR LF>
Instrument response: [CC]OK<CR LF> if the command has been received; the instrument’s response does not
mean necessarily that the instrument executes the tare.
Zero Command
[CC]ZERO<CR LF> or [CC]Z<CR LF>
Instrument response: [CC]OK<CR LF> if the command has been received; the instrument’s response does not
mean necessarily that the instrument executes the zero.
Clear Command
[CC]CLEAR<CR LF> or [CC]C<CR LF>
Instrument answer: [CC]OK<CR LF> if the command has been received, no answer for the
on/off command; the
instrument answer does not imply that the command is executed.
The command works also inside the setup mode.
Note