SCT-1100 Weight Transmitter
To Use Input/Output Weighing Mode
1. Press
to acquire first weight.
displays.
2. Press
again to acquire second weight.
displays. Data prints after acquisition of the second
weight.
Press and hold
to enter the setpoint input menu . See
.
The weighing cycle can be interrupted by pressing
after the acquisition of the first weight.
displays.
Press
to confirm the cancellation or another key to continue.
A weight is acquired when:
* The weight is stable and greater than zero
*The
parameter conditions have been met. See
.
3.10.4 Multi-Scale Repeater
A system consists of one or more (up to 32) primary instruments connected to one or more weighing systems. All
communicate with a secondary instrument, which acts as a weight repeater, on which it is possible to view or print
the weight of each individual scale or the sum of the weight detected by the single scales.
primary instruments repeat all the keys pressed on the secondary instrument.
A network can be configured with a secondary instrument which remotely performs the connected scale
functions and one or more listen only secondary instruments which only repeat the weight.
Figure 3-2. Multi-scale Repeater Flow Chart
Secondary Instrument Configuration
1. Turn on the instrument and press
while the firmware version displays.
displays.
2. Press
to enter the menu.
3. Press
or
until
displays. Press
to enter the menu.
4. Press
or
until
is displays. Press
to enter the menu.
5. Press once.
displays, then a prompt to enter a number.
6. Enter a number between 01 and 32 using
to select the digit to be modified and
or
to increase or
decrease the digit.
7. Press
to
confirm.
displayed.
Note
C
Note