NOTE:
The ability to edit the settings of the Timer tab varies by item.
Menu Idle Timeout
– Edit on the CM. Cannot edit through the RCU.
Amount of idle time before the RAMP® 200 returns to the home screen. To disable, enter
0
. Range: 0 -
10 minutes. Default: 5 minutes.
Update/Upload Start Date/Time
– Edit through the RCU only.
The date and time used to calculate the start of the following intervals:
1.
Update Interval
– Edit through the RCU only.
Interval at which the Reader checks for changes to settings in a networked environment. Disable by
entering
0
. Range: 0 - 1440 minutes. Default: 1.
2.
Upload Results Interval
– Edit through the RCU only.
Interval at which the Reader sends additions to the result and event logs in a networked environment. To
disable, enter
0
. Range: 0 - 1440 minutes. Default: 0 (Disabled).
NOTE:
Only new results not previously transferred are transferred during an upload. Data is not
deleted from the CM upon transfer. When the CM database is full, the oldest result is over-written.
3.
Auto IQC Interval
– Edit on CM and through the RCU.
The interval between automatic IQC self-diagnostics in either local or networked environment. To
disable, enter
0
. Range: 0 - 1440 minutes. Default: 0 (Disabled).
You must use the RCU to edit the Printer tab settings. They cannot be edited in the CM.
Automatic printing can be enabled independently for Assay, LQC and IQC results.
On
:
Prints the result automatically when the Assay is done.
Off
:
Does not print the result automatically. -
Default: Off
Summary of Contents for RAMP 200
Page 7: ...A ...
Page 13: ...B ...
Page 15: ...C ...
Page 20: ...Figure 4 RAMP 200 Test Module rear panel label ...
Page 21: ...D ...
Page 25: ...E ...
Page 31: ...F ...
Page 37: ...G ...
Page 44: ...H ...
Page 54: ......
Page 55: ...I ...
Page 62: ...J ...
Page 65: ...K ...
Page 73: ...L ...
Page 85: ...M ...
Page 94: ......