The RAMP® 200 can store up to 900 results:
300 patient results, maximum
300 LQC results, maximum
300 IQC results, maximum
When the CM database is full, the oldest result is overwritten. Dual analyte Assay results are stored as
two separate records. IQC results are stored separately for each individual Test port.
In manual transfer mode only (i.e., when the instrument is not connected to a network), the warning
‘
Database Almost Full, Backup Data’
is displayed when there are less than 20 LQC or patient result
spaces left in the database. This message will continue to be displayed until the database has cleared.
When the instrument is in auto transfer mode, only new results that have not previously been uploaded
are transferred during the upload. Data is not deleted from the CM upon transfer.
NOTE
Prior to deleting results, transfer and store all data in accordance with
applicable local, state and/or federal guidelines or accreditation
requirement and each facility’s protocol.
1
From the
Main Menu
screen (
Figure 6
), select
Results
. A list of all
Assay Results
is
displayed (
Figure 15
) in reverse chronological order, with the most recent results at the top.
Figure 15
Assay Results Screen