You must use the RCU to edit the LQC tab settings. They cannot be edited in the CM.
Default: Timer Reset (hours) = 0, LQC will not time-out.
To
Hide Numeric LQC Results
and report only PASS/FAIL.
Set as
Yes
or
No
.
Yes: Pass/Fail appears in place of results.
No: Actual result appears.
Default: No.
If LQC is required, display
LQC Timeout Warning (% remaining)
:
Before LQC expires (for any Assay), the LQC warning icon appears when the user-defined
LQC Timeout
Warning
counts down to its specified percentage. Default: 10%.
If LQC is required, display
Max Failures Before Lockout
:
The number of LQC failures allowed before Assay is locked out. Can be set from 1-9. Default: 1.
String length can be set from 1 to 18.
Default: Min Length = 1, Max Length = 18.
Use this tab to view the time and timer interval settings.