Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
7-27
7.10 Performance
Verification Data
Form
A performance verification consists of all the electrical safety, alarm/LED, and
pressure tests.
Complete this form at every performance verification. Make copies of this form
for data collection.
BiPAP Focus Performance Verification Data Form
Ventilator serial no.
Service notification
(Respironics use only)
Model no.
Total operating time
Hospital name and
address
Technician name
Date
Preliminary cleaning and inspection
Circle one
Damage to ventilator exterior?
Yes
No
If yes, brief description:
Cleaned ventilator exterior
Yes
No
Cleaned ventilator interior
Yes
No
Inspected inlet filters
Yes
No
Electrical safety test
Passed value
Failed value
Value after
repair
Forward leakage current:
< 100 microamperes (100-120 VAC)
< 300 microamperes (230-240 VAC)
uA
uA
uA
Reverse leakage current:
< 300 microamperes (100-120 VAC)
< 500 microamperes (230-240 VAC)
uA
uA
uA