Chapter 5
Diagnostic Mode and Troubleshooting
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
5-17
Use the arrow keys to scroll through the log.
lists event IDs and
corrective actions.
NOTE:
When performing corrective actions:
•
Perform the corrective actions in the order listed.
•
Always retrieve and save the DRPT if possible
before
replacing
parts. If the corrective action suggests swapping the user
interface or controller PCB, retrieve the DRPT
before
installing
the replacement PCB.
•
To verify that replacing a part corrects the problem,
temporarily install (swap) a replacement part. If the
replacement corrects the problem, verify that the problem
recurs with the original part before replacing it.
Table 5-3: Significant Event Codes
Event ID
Comment / Corrective Action
1
GENERIC SOFTWARE ERROR
Controller processor detected a software error.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
2
SOFTWARE CORRUPT
Controller processor detected a program memory checksum error.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
3
EXTERNAL RAM FAILURE
Controller processor detected an external RAM memory error.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
4
FIQ STACK OVERFLOW
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
5
FIQ STACK UNDERFLOW
Controller processor detected a stack underflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
6
NESTED IRQ STACK
OVERFLOW
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.
7
NESTED IRQ STACK
UNDERFLOW
Controller processor detected a stack underflow.
1. Retrieve the DRPT and contact technical support.
8
IRQ STACK OVERFLOW
Controller processor detected a stack overflow.
1. If the problem persists replace the controller PCB.
2. Retrieve the DRPT and contact technical support.