Chapter 4
Periodic Maintenance
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
4-5
4.4.3
Battery Replacement
Follow these steps to replace the battery (Figure 4-2):
1. Turn the system
OFF (O)
and disconnect the DC power supply.
2. Remove the two screws that secure the battery compartment door.
3. Remove the old battery: remove from the compartment and unplug the
cable.
4. Install the new battery: plug in the cable and install the battery into
the compartment.
5. Replace the battery compartment door and screws.
6. Reconnect the DC power supply.
7. Allow sufficient time to fully charge the new battery (Battery is fully
charged when the
Battery Charging
LED turns off).
Figure 4-2: Replacing the Battery
CAUTION:
Use only Respironics-approved batteries.
Battery compartment
Battery cable
connector
Battery
compartme
nt door