Chapter 10
10-10
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
REF 1029568 Rev B
Specifications
Guidance and manufacturer’s declaration: electromagnetic immunity, IEC
60601-1-2 + A1:2000.
The BiPAP Focus System is suitable for use in the specified electromagnetic
environment. The customer and/or the user of the BiPAP Focus System should
ensure that it is used in an electromagnetic environment as described below:
RF emissions CISPR 11/Class B
The BiPAP Focus System is suitable
for use in all establishments other
than domestic and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2/
(not applicable)
Voltage fluctuations/flicker emissions
IEC 61000-3-3/ (not applicable)
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
Mains power quality should be that
of a typical commercial and/or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV
differential
mode
± 2 kV
common
mode
± 1 kV
differential
mode
± 2 kV
common
mode
Mains power quality should be that
of a typical commercial and/or
hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
Emissions Test/Compliance
Electromagnetic Environment
Guidance