Chapter 8
Component Removal/Installation
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
8-11
Figure 8-12: Controller PCB connection points
6. Disconnect as needed:
•
Flow valve cable
•
Switch control cable
•
Backup alarm cable
•
Blower cable
•
Interprocessor communications cable
•
Interconnect power cable
•
Tubing from the pressure transducers (MT2-P1, MT2-P2, MT3,
MT4): label tubing to ensure correct reassembly.
NOTE:
To avoid damaging the controller PCB pressure transducers, carefully
remove tubing to avoid pulling on the sensors.
NOTE:
Switch control
cable
Blower cable
J5
Backup alarm
cable
MT2-P2 connects
to air flow sensor
nipple (away from
patient outlet)
Flow valve J9
Interconnect
power cable J7
Interprocessor
communications
cable J1 and J10
MT3 blower
sensor port
MT4 patient
outlet port
MT2-P1 connects
to air flow sensor
nipple (closest to
patient outlet)