Chapter 1
Introduction and Intended Use
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
1-3
1.2 Where to Go for
Help
For customer service and product support contact:
Respironics, Inc.
USA telephone: 1-800-345-6443
Fax: 1-800-886-0245
International telephone: 724-387-4000
International fax: 724-387-5012
[email protected].
[email protected].
www.respironics.com.
Pliers
Local supplier
Pliers, needle nose
Local supplier
Screwdriver, flat, #2
Local supplier
Screwdriver, flat, pen size
Local supplier
Screwdriver, Phillips, #1
Local supplier
Screwdriver, Phillips, #2
Local supplier
Screwdriver, Phillips, pen size
Local supplier
Torque driver (1 to 25 in.-lbs)
Local supplier
Tubing, silicone, 3/16-in. ID x 6.5 ft.
Local supplier
Table 1-2: Measurement Accuracy Requirements
Unit of
Measure
Range
Accuracy
Pressure
-25 to 150 cmH20
0-100 psi
+/- 1% of reading or +/- 0.20 cmH20 (whichever is
greater).
Flow (standard)
0-300 SLPM
+/- 2% of r/- 0.20 SLPM (whichever is greater)
Torque
5 to 25 in-lbs.
+/- 1 in-lbs. of reading
Voltage
DC: +/- 5 to 50V
AC: 2 to 300 V
+/- 2% of reading
Table 1-1: Test Equipment, Tools, and Materials (Continued)
Description
Manufacturer and Model