Chapter 7
Testing, Calibration, and Performance Verification
REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
7-21
27. Adjust the flow control valve until 10 LPM is displayed on the
flowmeter, then click
OK
As the program calibrates the negative portion of the flow table, the
flow meter may not display a negative flow. However, pressure against
the normal flow direction is considered a negative flow.
Figure 7-22: Negative Flow Value Display
28. Enter the flow reading (Figure 7-23). Be sure to enter the negative or
minus symbol first. The software rejects a positive value at this point.
Figure 7-23: Entering The Negative Flow Value
29. At the prompts, adjust the flow control valve to achieve various
negative flows, enter the flowmeter readings, then click
OK
.
30. At the
Remove Negative Flow Resource
prompt, remove the negative
flow source and click
OK
.