REF 1029568 Rev B
BiPAP Focus Ventilator Service Manual © Respironics, Inc.
8-1
Chapter 8. Component Removal/Installation
This chapter describes how to remove and install replaceable ventilator
components.
WARNING:
CAUTION:
NOTE:
Always retrieve and save the DRPT if possible
before
replacing parts.
Removing the user interface PCB deletes the serial number, significant
event log entries, and ventilator preference settings. Review the
significant event log before removing the user interface PCB.
NOTE:
To avoid damaging the controller PCB pressure transducers, carefully
remove tubing to avoid pulling on the sensors.
NOTE:
When reconnecting the tubing to the controller PCB sensors, ensure that
the tubing fits tightly. If the tubing has become loose, clip a short length
(2 mm) from the end of the tubing to ensure a snug fit.