3156B User Manual
Maintenance and Performance Checks 6-32
Racal Instrument © 2005
7. Configure the 3156B channel 1 as follows:
Reset
Pattern
Rate: 10Mpps
Wave Mode: Digital
Digital Mode: Freerun
Enable:
On
8. Using WaveCAD, open the Digital Pattern Stim List Table,
prepare and download the following table:
Index
Stim List (hex)
Hold Count
1
1
10000000
2
2
10000000
3
4
10000000
4
8
10000000
5
10
10000000
6
20
10000000
7
40
10000000
8
80
10000000
9
100
10000000
10
200
10000000
11
400
10000000
12
800
10000000
Test Procedure
1. Watch the LED’s on the test board. The 12 MSD LED’s light in
sequence. Each LED lights for about 1 second
2. Set up the oscilloscope and check the output level. Verify output
level is ECL
Test Results
Pass
Fail
3. Repeat step 8 but change hold count to 100 for the entire list
4. Set up the oscilloscope and verify digital sequence of 10
μ
s per
pattern step
Test Results
Pass
Fail
5. Turn chassis power off and remove the cable from 3156B
channel 1
6. Repeat the test procedure as above for channel 2
Test Results
Pass