3156B User Manual
Racal Instrument © 2005
Maintenance and Performance Checks 6-31
Test Procedure
1. Verify AM operation on the oscilloscope as follows:
Waveform:
Amplitude Modulated Sine
Modulation Depth: 10kHz
Test Results
Pass
Fail
2. Remove the cable from 3156B channel 1 and connect to chan 2
3. Repeat the test procedure as above for channel 2
Test Results
Pass
Fail
Digital Pattern
Generator
Characteristics
Digital Patterns -
Freerun
Equipment
: Oscilloscope, Digital pattern test board, Dual DC power
supply
Preparation
:
1. Turn chassis power OFF
2. C5V and -5V to the test board
3. Hook up the test board on the channel 1 digital output connector
4. Turn chassis power ON
5. Turn power supply power ON
NOTE
Do not attempt to connect the test board to the 3156B
connector while power is ON as this may result in
permanent damage to the 3156B. Always turn chassis
power OFF before connecting or disconnecting the test
board to the 3156B.
6. Connect 3156B Channel 1 output to the oscilloscope input,
channel 1