3156B User Manual
Maintenance and Performance Checks 6-28
Racal Instrument © 2005
Test Results
Pass
Fail
2. Remove the cable from 3156B channel 1 and connect to chan 2
3. Repeat the test procedure as above for channel 2
Test Results
Pass
Fail
Variable Dwell Time
Frequency Hops
Equipment
: Oscilloscope (LeCroy LT342, fitted with jitter package)
Preparation
:
1. Configure the oscilloscope as follows:
Time Base:
0.1 ms
Sampling Rate: 50MS/s at least.
Trace A View: Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1V/div
2. Connect 3156B Channel 1 output to the oscilloscope input,
channel 1
3. Connect the 3156B SYNC output to the oscilloscope input,
channel 2
4. Configure the 3156B channels 1/2 as follows:
Reset
Wave Mode: Modulation
Modulation Mode: HOP
Hop
Mode:
Variable
Sync:
On
Output:
On
Enable:
On
5. Using WaveCAD, open the Hop Table composer and download
the following table (both channels):
Frequency Dwell
Time
1.0e6 50e-6
1.2e6
100e-6
1.4e6
150e-6
1.6e6
200e-6
1.8e6
250e-6
2.0e6
300e-6
2.2e6
350e-6
2.4e6
400e-6
2.6e6
450e-6
2.8e6
500e-6