3156B User Manual
Racal Instrument © 2005
Maintenance and Performance Checks 6-13
4. Connect the function generator output to the 3156B TRIG IN
connector
5. Configure the 3156B channels 1/2 as follows:
Reset
Frequency: 1.5625MHz
Waveform: Sinewave
Burst Count:
1e6 counts, each channel
Amplitude: 5V
Enable
Source:
Hardware
Output:
On
Enable:
On
Test Procedure
1. Perform trigger and gate tests using Tables 5-13
2. Configure the counter to TOTB Measurements and perform burst
tests using Tables 5-13
Table 6-13 Trigger, gate, and burst Characteristics
3156B Run
External Trigger
Oscilloscope Reading
Mode
Pulse
CH 1
CH 2
Pass
Fail
Triggered
10kHz, Continuous
Triggered waveform
Triggered waveform
Gated 10kHz,
Continuous
Gated
Waveform Gated
Waveform
Burst
10kHz, Single shot
Burst, 106 waveforms Burst, 106 waveforms
Delayed Trigger
Equipment
: Function generator, 50
Ω
“T” connector, Counter, Wave
CAD
Preparation
:
1. Configure the Function generator as follows:
Amplitude: 1V
Frequency:
1MHz
Trigger Mode: Triggered.
Wave:
Squarewave
2. Place the “T” connector on the output terminal of the function
generator. Connect one side of the “T” to the 3156B TRIG IN
connector and the other side of the “T” to the channel A input of
the counter
3. Connect the 3156B output to channel B input of the counter
4. Configure the counter to TI A to B measurements
5. Using WaveCAD prepare and download the following waveform:
Wavelength: 100
points
Waveform:
Pulse, Delay = 0.1, Rise/Fall = 0,
High Time = 99.99
6. Configure the 3156B channels 1/2 as follows:
Reset
SCLK:
200MHz
Waveform: Arbitrary
Run
Mode: Trigger