Maintenance and Performance Checks 6-10
Racal Instrument © 2005
Sinewave
Spectral Purity ,
DAC Waveforms
Equipment
: Spectrum Analyzer
Preparation
:
1. Connect 3156B Channel 1/2 output to the spectrum analyzer
input. Use 50
Ω
and 20dB feedthrough termination at the
spectrum analyzer input
2. Configure the 3156B channels 1/2 as follows:
Reset
Frequency:
As required by the test
Waveform: Sinewave
Amplitude: 5V
Output:
On
Enable:
On
Test Procedure
1. Perform sinewave spectral purity, DAC waveforms tests using
Table 5-9
Table 6-9 Sinewave Spectral Purity, DAC Waveforms Test
3156B Freq
Reading
Spectrum Analyzer, Settings & Results
Settings
Limits
Start
Stop
CH 1
CH 2
Pass
Fail
1 MHz
>55 dBc
100K
10M
5 MHz
>40 dBc
1M
20M
10 MHz
>35 dBc
1M
100M
25 MHz
>30 dBc
1M
150M
Sinewave
Spectral Purity,
Modulation
Waveforms
Equipment
: Spectrum Analyzer
Preparation
:
1. Connect 3156B Channel 1/2 output to the spectrum analyzer
input. Use 50
Ω
and 20dB feedthrough termination at the
spectrum analyzer input
2. Configure the 3156B channels 1/2 as follows:
Reset
Waveform Mode: MOD
Modulation Mode: OFF
CW
Frequency:
As required by the test
Amplitude
5V
Output:
On
Test Procedure
1. Perform sinewave spectral purity, DDS Waveforms tests on both
channels using Table 5-10