![background image](http://html2.mh-extra.com/html/racal-instruments/1269/1269_user-manual_3121614091.webp)
1269 User Manual
Using The Enhanced Monitoring System 4-22
TTLT7
NONE
The Default state for this command is “NONE”.
Connecting The TRIG
OUT Output To a
TTLTRG Line
The TRIG OUT output may be connected to one of the
TTLTRGx lines (TTLTRG0 through TTLTRG7) . The following
commands may be used to connect the TRIG OUT output to
one of the TTLTRG lines, or to disconnect the input from all of
the TTLTRG lines:
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG0
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG1
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG2
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG3
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG4
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG5
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG6
VXI:CONF:MONITOR:TRIG:OUTPUT TTLTRG7
VXI:CONF:MONITOR:TRIG:OUTPUT NONE
VXI:CONF:MONITOR:TRIG:OUTPUT INPUT
The first eight commands above connect the TRIG OUT output
to TTLTRG0 through TTLTRG7, respectively. The ninth
command disconnects the TRIG OUT output from all of the
TTLTRG lines. The last command connects the TRIG IN input
to the TRIG OUT output without affecting any of the TTLTRG
lines. In this case, if the TRIG IN input is already connected to
a TTLTRG line, it will be disconnected prior to connecting it to
the TRIG OUT output.
Note that by selecting one TTLTRG line, all other TTLTRG
lines are implicitly disconnected.
To determine which line is presently connected, the query:
VXI:CONF:MONITOR:TRIG:OUTPUT?
may be used. The reply to this query will be one of the following:
TTLT0
TTLT1
TTLT2
TTLT3
TTLT4
TTLT5
TTLT6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com