1269 User Manual
Using The Enhanced Monitoring System 4-44
busy writing data for more than 40 milliseconds. This
indicates a general EEPROM failure. This error may be
returned if the *TST? command returns a value of -5.
Time-of-day real-time clock failed
This error is returned when the real-time clock on the EMS
does not advance after a minimum of 1.2 seconds. This
error may be returned if the *TST? command returns a
value of -6.
Power-on time real-time clock failed
This error indicates that the power-on time real-time clock
on the EMS does not advance after a minimum of 1.2
seconds. This error may be returned if the *TST?
command returns a value of -6.
The following self-tests are NOT performed when VXI
mainframe power is NOT applied:
EEPROM tests
real-time clock tests