1269 User Manual
Error Messages B-3
Table B-1, Error Messages (Continued)
Error Message
ESR Bit Set
Error Description
-144,”Character data too long”
CME
The character data element contains more than 12
characters.
-148,”Character data not allowed”
CME
A legal character data element was received but is not
allowed.
-150,”String data error”
CME
A string received as a command parameter is invalid.
-151,”Invalid string data”
CME
A string was received where expected, but contained an
invalid character (e.g. was not terminated with a quote
character).
-158,”String data not allowed”
CME
A string data element was encountered but is not allowed.
-200,”Execution error”
EXE
An execution error has occurred. The EMS cannot determine
a more specific problem with the error (see errors -201 to -299
below).
-210,”Trigger error”
EXE
A trigger error has occurred. No more specific info is known.
-211,”Trigger ignored”
EXE
*TRG or GET received but ignored due to timing
considerations.
-212,”Arm ignored”
EXE
An arming signal was received and recognized but was
ignored.
-213,”Init ignored”
EXE
A request for measurement initiation was ignored since
another measurement was already.
-214,”Trigger deadlock”
EXE
Trigger subsystem is deadlocked.
-215,”Arm deadlock”
EXE
Trigger arming is deadlocked.
-220,”Parameter error”
EXE
Command parameter is invalid or unrecognized.
-221,”Settings conflict”
EXE
Legal command parameter received but could not be
executed because doing so would place the device in an
invalid state.
-222,”Data out of range”
EXE
Command data parameter is out of range.
-223,”Too much data”
EXE
Block, expression, or string type received contained too much
data.
-224,”Illegal parameter value”
EXE
The parameter value is not from list of possible values.
-240,”Hardware error”
EXE
An (unspecified) hardware error prevented proper execution
of the command.
-241,”Hardware missing”
EXE
The command failed because a hardware option was not
installed.
-300,”Device-specific error”
DDE
A generic device-specific error which does not match one of
the SCPI-defined errors in the range -300 to -399 has
occurred.
-310,”System error”
DDE
A device-defined generic system error has occurred.
-311,”Memory error”
DDE
The EMS’s memory is corrupted.