1269 User Manual
Using The Enhanced Monitoring System 4-12
Table 4-3, Temperature Condition and Event Register Bit Assignments
Bit
Set When
0
VXI slot C0 temperature rise exceeds limit
1
VXI slot C1 temperature rise exceeds limit
2
VXI slot C2 temperature rise exceeds limit
3
VXI slot C3 temperature rise exceeds limit
4
VXI slot C4 temperature rise exceeds limit
5
VXI slot C5 temperature rise exceeds limit
6
VXI slot B0 temperature rise exceeds limit
7
VXI slot B1 temperature rise exceeds limit
8
VXI slot B2 temperature rise exceeds limit
13
Ambient temperature exceeds limit
Table 4-4, Fan Speed Condition and Event Register Bit Assignments
Bit
Set When
0
Fan is out of tolerance
Table 4-5, Time Condition and Event Register Bit Assignments
Bit
Set When
0
Present Power-On Time exceeds limit
1
Cumulative Power-On Time exceeds limit
2
Filter Service Time exceeds limit