Specifications A-8
Table A-8, EMS Monitoring System Specifications, ( Typical )
Characteristics
Description
Auxiliary Rear Panel DC Voltage Outputs
Voltage Current
+5 V 1 Amp maximum, (Self Healing
Fuse)
+12 V 1 Amp maximum, (Self Healing
Fuse)
+24 V 1 Amp maximum, (Self Healing
Fuse)
Maximum slot temperature (
Tmax
)
Rear
Panel Output Voltage
100 mV per
o
C, 0
o
C corresponds to 0 V output
± 1.5
o
C accuracy,
-5 to + 85
o
C range
Maximum slot Temperature Rise (
)
)
T)
Rear
Panel Output Voltage
100 mV per
o
C, 0
o
C corresponds to 0 V output
± 1.5
o
C accuracy,
-5 to + 85
o
C range
Fan Speed
TACH
Rear Panel Output
2 Pulses/Rev, TTL level Tach output
Fan Speed Range, typical
2000 to 3400 RPM
VXI Power Supply
Voltage
Health Status
Readout
Available at :
Front Panel Display,
VXI Message Based Interface,
RS-232 interface
Default Tolerance limits: V nom -8.75%,
+10%
3% accuracy
VXI Power Supply
Temperature
Health
Status Readout
Available at :
Front Panel Display (1 x 10 Characters),
VXI Message Based Interface,
RS-232 interface
Default Tolerance limits:
)
)
T
> 20
o
C,
Tamb
>
70
o
C
± 1.5
o
C accuracy,
-5 to + 85
o
C range
VXI Power Supply
Airflow
Health Status
Readout
Available at :
Front Panel Display (1 x 10 Characters),
VXI Message Based Interface,
RS-232 interface
Tolerance limit: 2000 RPM < Fan speed <
5000 RPM
± 200 RPM accuracy