1269 User Manual
Configuring The 1269 2-27
Installing The
VXI
plug&play
Driver
After the EMS has been installed into the 1269 mainframe, the
VXI
plug&play
software may be used to communicate with the
EMS. To install the software, first connect the host computer
power on the mainframe (see
Powering On The Mainframe
in
Chapter 3
), then perform the following operations:
1.
Start Windows (3.1 or 3.11) on your computer if it is not
already running.
2.
Insert the VXI
plug&play
installation disk #1 into the 3-
1/2" floppy disk drive.
3.
Run the SETUP program on the installation disk.
4.
Follow the instructions presented by the SETUP
program
After the SETUP program has completed, the executable Soft
Front Panel program may be run. To run the Soft Front Panel,
ensure that the following conditions are met:
1.
The computer is connected to the 1269 VXI mainframe
via a MXI/VXI interface, a GPIB/VXI interface, or the
computer is an embedded VXI computer.
2.
VISA is loaded onto your system. VISA is a library of
functions which provide communication between a
computer and instruments (GPIB and VXI). VISA may
be obtained from the manufacturer of the MXI/VXI,
GPIB/VXI, or embedded computer.
3.
The 1269 mainframe has power applied and the power
switch has been turned ON.
4.
For the MXI/VXI and embedded computers, the resource
manager program has been run since the 1269
mainframe power was last turned ON.
To run the Soft Front Panel, “double-click” on the 1269 EMS
Front Panel” icon in the “VXIPNP” Windows Group.
If the four conditions above are met, the Soft Front Panel
program will automatically locate the EMS in the 1269
Mainframe. The Soft Front Panel program will display the VXI
logical address of the EMS, and the “Active” LED on the Soft
Front Panel will be green.