![background image](http://html2.mh-extra.com/html/racal-instruments/1269/1269_user-manual_3121614123.webp)
1269 User Manual
SCPI Command Reference 5-10
Command
Parameter Limits
The parameter limits for the numeric commands are shown in
Table 5-2.
Table 5-2, Defaults, Minima and Maxima
Command
Default
Minimum
Maximum
Notes
SENSE:VOLT1:RANGE:UPPER
5.4
5.0
40.0
+5V PS voltage upper tolerance
SENSE:VOLT1:RANGE:LOWER
4.6
0.0
5.0
+5V PS voltage lower tolerance
SENSE:VOLT2:RANGE:UPPER
-4.8
-5.2
0.0
-5.2V PS voltage upper tolerance
SENSE:VOLT2:RANGE:LOWER
-5.6
-40.0
-5.2
-5.2V PS voltage lower tolerance
SENSE:VOLT3:RANGE:UPPER
-1.8
-2.0
0.0
-2V PS voltage upper tolerance
SENSE:VOLT3:RANGE:LOWER
-2.2
-16.0
-2.0
-2V PS voltage lower tolerance
SENSE:VOLT4:RANGE:UPPER
25.9
24.0
100.0
+24V PS voltage upper tolerance
SENSE:VOLT4:RANGE:LOWER
22.1
0.0
24.0
+24V PS voltage lower tolerance
SENSE:VOLT5:RANGE:UPPER
-22.1
-24.0
0.0
-24V PS voltage upper tolerance
SENSE:VOLT5:RANGE:LOWER
-25.9
-100.0
-24.0
-24V PS voltage lower tolerance
SENSE:VOLT6:RANGE:UPPER
12.9
12.0
100.0
+12V PS voltage upper tolerance
SENSE:VOLT6:RANGE:LOWER
11.1
0.0
12.0
+12V PS voltage lower tolerance
SENSE:VOLT7:RANGE:UPPER
-11.1
-12.0
0.0
-12V PS voltage upper tolerance
SENSE:VOLT7:RANGE:LOWER
-12.9
-100.0
-12.0
-12V PS voltage lower tolerance
SENSE:FREQ:RANGE:UPPER
5200.0
2000.0
7650.0
Fan, upper RPM tolerance
SENSE:FREQ:RANGE:LOWER
2000.0
500.0
7650.0
Fan, lower RPM tolerance
SENSE:TEMP14:RANGE:UPPER
70.0
0.0
140.0
Ambient temperature upper tol.
SENSE:TEMP<X>:RANGE:UPPER
20.0
0.0
140.0
VXI slot temp rise tol.
<X> = 1 to 9
SENSE:TIME1:RANGE:UPPER
31536000
(365 days)
0.0
125 years
Power-On elapsed time limit
SENSE:TIME2:RANGE:UPPER
157680000
(5 x 365 days)
0.0
125 years
Cumulative power-on time limit
SENSE:TIME3:RANGE:UPPER
2592000
(30 days)
0.0
125 years
Filter service power-on time limit
SENSE:VXI:IACK<X>:LIMIT
0
0
255
IACKn limit (<X> = 1 to 7)
SENSE:VXI:BERR:LIMIT
0
0
255
VXI Bus Error limit
VXI:CONF:MON:TRIG:DELAY:TIME
0.0
0.0
1048575968.75
(nanoseconds)
Trigger Delay time
resolution = 31.25 nanoseconds
Device Specific
Commands
The 1269 EMS conforms to the 1994.0 version of the SCPI
standard. Some of the commands used are not included in the
1994.0 version. However, these commands are designed with
the SCPI standard in mind and they follow all of the command
syntax rules defined by the standard.
Table 5-1
lists all device-
specific commands that were designed specifically for the EMS
as “Not Confirmed” to the 1994.0 SCPI version.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com