1269 User Manual
Using The Enhanced Monitoring System 4-21
TRIG IN signal and the TTLTRGx line or the
TTLTRGx signal and the TRIG OUT line.
4)
Route the TRIG IN front panel input signal directly
to the TRIG OUT front panel output (with or
without a delay).
Note that by routing the TRIG IN input to the same TTLTRGx
line that is connected to the TRIG OUT output, the TRIG IN
signal may be connected to the TRIG OUT connector in
addition to the TTLTRGx line.
Connecting The TRIG
IN Input To a TTLTRG
Line
The TRIG IN input may be connected to one of the TTLTRGx
lines (TTLTRG0 through TTLTRG7) . The following commands
may be used to connect the TRIG IN input to one of the
TTLTRG lines, or to disconnect the input from all of the
TTLTRG lines:
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG0
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG1
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG2
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG3
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG4
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG5
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG6
VXI:CONF:MONITOR:TRIG:INPUT TTLTRG7
VXI:CONF:MONITOR:TRIG:INPUT NONE
The first eight commands above connect the TRIG IN input to
TTLTRG0 through TTLTRG7 respectively. The last command
disconnects the TRIG IN input from all of the TTLTRG lines.
Note that by selecting one TTLTRG line, all other TTLTRG
lines are implicitly disconnected.
To determine which line is presently connected, the query:
VXI:CONF:MONITOR:TRIG:INPUT?
may be used. The reply to this query will be one of the following:
TTLT0
TTLT1
TTLT2
TTLT3
TTLT4
TTLT5
TTLT6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com