1260-00C User Manual
Local Command Set 4-40
•
*IDN?
•
*OPC
•
*OPC?
•
*RST
•
*SRE
•
*SRE?
•
*STB?
•
*TRG
•
*TST?
•
*WAI
These commands conform to the minimal 488.2 requirements.
Many of these 488.2 commands have limited meaning in the VXI
environment, but are included for compatibility. The 1260-00C has
no reason to interrupt as a 488.2 instrument. It is only a parser.
All other functions of the 1260-00C are considered to be interface
functions for other 488.2 VXI devices. It is the responsibility of
each VXI device connected via the GPIB to the 1260-00C to
implement 488.2 protocols if individual device 488.2 compatibility
is required.
*CLS
488.2
Intent:
Clear the device status data structures, and force it
to the Operation Complete Query Idle state.
Command
Syntax:
*CLS
Action: None
*ESE
488.2
Intent:
Set the 1260-00C's Standard Event Status
Enable (ESE) register bits.
Command
Syntax:
*ESE <byte value>