1260-00C User Manual
Local Command Set 4-2
Command Set
Access
Execute the local commands from the following ports:
RS-232
GPIB
VXI Word Serial Communication
Individual Code Instruments
All ports are active when the 1260-00C is in the Normal Operation
substate, and operate independently of one another. The
1260-00C returns query responses only to the port originating the
query. The 1260-00C also maintains a separate status state for
each port. Use local commands to disable and re-enable each
port's access to the local command set. The RS-232 port prompts
you to enter a local command with the
1260-00C>
prompt.
Command Syntax
The local command set parser is syntactically compatible with the
IEEE-488.2 standard. It will accept numeric parameters in the
488.2 binary, octal, decimal, or hexadecimal formats. 488.2 binary
parameters are prefixed with #b. Octal parameters are prefixed
with #q, and hexadecimal parameters are prefixed with #h. Table
4-1 lists the most common numeric parameter types. The ranges
given in Table 4-1 apply unless otherwise specified.
Table 4-1. Valid Ranges For Common Numeric Command Parameters
Parameter
488.2 Decimal
488.2 Hexadecimal
<logical address>
0 to 254
#h0 to #hFE
<GPIB address>
0 to 30
#h0 to #h1E
<handler>
1 to 3
#h1 to #h3
<level>
0 to 7
#h0 to #h7
<A16 address>
0 to 65535
#h0 to #hFFFF
<A24 address>
2097152 to 14680062
#h200000 to #hE7FFFE
<word value>
0 to 65535
#h0 to #hFFFF
<byte value>
0 to 255
#h0 to #hFF
<Boolean>
0 to 1
#h0 or #h1
The logical value of a
<Boolean>
parameter is FALSE for the