1260-00C User Manual
Operation 3-14
Table 3-8, CLK10 Routing Options
Switch S15
Switch S59
Function
OFF
OFF
CLK10 sourced from onboard clock.
ON
OFF
CLK10 and EXT CLK connector sourced from onboard clock.
OFF
ON
CLK10 sourced from an external clock via the EXT CLK
connector.
ON
ON
Invalid. Do not use this setting.
Slot 0 Resource Manager
Operation
At start-up, a 1260-00C configured as a Slot 0 Resource Manager
performs its self-tests, executes the RM functions, and enters its
normal mode of operation.
Front Panel LED
Indications for RM
Operation
The five front panel LEDs are SYSFAIL, FAILED, TEST, ON LINE,
and ACCESS. The 1260-00C uses the FAILED, TEST, and ON
LINE LEDs to indicate the progress of its self-initialization,
self-test, and RM functions. The LED indications are shown in
Table 3-9. A successful system start-up will sequence through the
first five states. The point of failure is indicated for states in which
the FAILED LED is lit for an extended period of time. The LED
indications are identical for Slot 0/Non-Slot 0 Resource Manager
operation.
Table 3-8. Front Panel LED Indications For RM Operation
Sequence
FAILED
TEST
ON LINE
State
Point of Failure
1
OFF
OFF
OFF
No power
Failed before self-
test
2
ON
OFF
OFF
In self-initialization
Failed in self-test
3
ON
ON
OFF
In self-test
4
OFF
ON
ON
Performing RM
5
OFF
OFF
ON
Online
ON
ON
ON
Failed
Failed while in RM
ON
OFF
ON
Failed
Failed while online
OFF
ON
OFF
In non-volatile
configuration, or
diagnostics mode