1260-00C User Manual
Operation 3-20
Table 3-11, Non-Slot 0 Resource Manager Operation Switch and Jumper Settings
Jumper/Switch
Position
Function
Switch S15
OFF
If S15 is ON, the 1260-00C
also routes CLK10 to the EXT
CLK connector on the front
panel.
Switch S22
OFF
VXI BTO disabled.
Switch S23
OFF
Bus arbiter and SYSCLK
disabled. CLK10 receiving
from backplane.
Switch S24
OFF
MODID pulled down.
Logical Address
See Chapter 5.
Logical address is 0. Set in
non-volatile configuration or
use the DIP switch.
Non-Slot 0 Model Code
See Chapter 5.
Model code is set to the
Non-Slot 0 value. Set in
non-volatile configuration.
Non-Slot 0 Resource
Manager Operation
The start-up sequence for a 1260-00C configured for Non-Slot 0
Resource Manager operation is nearly identical to the Slot 0
Resource Manager operation. In Non-Slot 0 RM operation, the
1260-00C controls the Slot 0 resources remotely.
A VXIbus Slot 0 device must be in the system. It must be either a
Register-Based device that implements the MODID Register, or a
Message-Based device that supports the Word Serial commands
Read MODID, Set Lower MODID, and Set Upper MODID. VXIbus
Specification Revision 1.2 Message-Based Slot 0 devices are not
supported.
Non-Slot 0
Message-Based
Device
Configuration (Non-
Resource Manager)
To configure the 1260-00C for Non-Slot 0 Message-Based
operation, use the following procedure:
a. Disable the VXIbus Slot 0 functions.
b. Set the model code of the 1260-00C to be configured for
Non-Slot 0 operation using the non-volatile configuration
mode.
c. Set the logical address to a non-zero value with an