![Racal Instruments 1260-00C User Manual Download Page 104](http://html1.mh-extra.com/html/racal-instruments/1260-00c/1260-00c_user-manual_3123646104.webp)
1260-00C User Manual
Local Command Set 4-55
to the value of
<line>
as follows:
Value of
<line>
Value of
<line text>
0 to 7
TTL <line>
8 to 9
ECL (<line> - 8)
40 to 49
GPIO (<line> - 40)
50
TCNTR
60
TICK1
and the meaning of
<protocol text>
corresponds to the value of
<protocol>
as
follows:
Value of
<protocol>
Value of
<protocol text>
0
ON
1
OFF
2
START
3
STOP
4
SYNC
5
SEMI-SYNC
6
ASYNC
7
SEMI-SYNC wait ACK
8
ASYNC wait ACK
ffffh
wait ACK ABORT
Example:
Source TTL line 4 for SEMI-SYNC protocol.
SrcTrig 4, 5, 0L
TrigAsstConf
Purpose:
Configure a specified TTL/ECL trigger line assertion
method.
Query
Syntax:
TrigAsstConf <line>, <mode>
where the value of
<line>
corresponds to the
trigger line to configure:
Value Trigger Line
0 to 7
TTL trigger lines 0 to 7
8 to 9
ECL trigger lines 0 to 1
ffffh
General assertion configuration (all lines)
and
<mode>
specifies the configuration mode,
where
Bit Specific Line Configuration Modes
0
1 = Synchronize falling edge of CLK10
0 = Synchronize rising edge of CLK10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com